CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

Cat# USCN-SCL231Hu-96

Size : 96T

Brand : USCN

Request more information

Contact local distributor :


Phone : +1 850 650 7790
CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

ZA2G; ZAG

  • Product No.SCL231Hu
  • Organism SpeciesHomo sapiens (Human) Same name, Different species.
  • Test MethodDouble-antibody Sandwich
  • Assay Length2h, 40min
  • Detection Range0.82-600ng/mL
  • SensitivityThe minimum detectable dose of this kit is typically less than 0.35ng/mL.
  • Sample TypeSerum, plasma, urine, saliva, seminal plasma, sweat and other biological fluids.
  • Download Instruction Manual
  • UOM 48T96T 96T*5 96T*10 96T*100
  • CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) Packages (Simulation)
  • CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) Packages (Simulation)
  • CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) Results demonstration
  • SCL231Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

This assay has high sensitivity and excellent specificity for detection of Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1).
No significant cross-reactivity or interference between Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) and analogues was observed.

Recovery of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

Matrices listed below were spiked with certain level of recombinant Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) and the recovery rates were calculated by comparing the measured value to the expected amount of Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-98 89
EDTA plasma(n=5) 85-104 92
heparin plasma(n=5) 84-105 98

Precision of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-105% 91-99% 84-103% 92-101%
EDTA plasma(n=5) 98-105% 88-101% 92-99% 80-91%
heparin plasma(n=5) 78-97% 98-105% 79-94% 87-103%

Stability of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1)

The microplate provided in this kit has been pre-coated with an antibody specific to Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Alpha-2-Glycoprotein 1, Zinc Binding (aZGP1) level in the sample or standard.;