Bovine Angiotensin II (Angiotensin II) ELISA Kit

Cat# EB0156

Size : 96T

Brand : FineTest

Request more information

Contact local distributor :


Phone : +1 850 650 7790

This kit is based on Competitive-ELISA detection method and takes 1h assay time. The microplate provided in this kit has been precoated with anti ANGII antibody. Sample or standard completely mixes with fixed amount of HRP labeled ANGII working solution and also add into each well. ANGII in the sample or standard competes with biotin-labeled ANGII to bind with antibody precoated on the microplate. Free components are washed off. Add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Compare OD450 value of sample with the standard curve through curve fitting software to determine the concentration of ANGII in the sample. The concentration of the target substance is inversely proportional to the OD450 value.

Catalogue No.: 
EB0156

Alias: 
ANG II ELISA Kit, Angiotensin II ELISA Kit

Species: 
Bovine

Range: 
1.563-100pg/ml

Sensitivity: 
0.938pg/ml

  • SPECIFICATIONS
  • FIGURES
  • CONDITIONS
Product Name
Bovine Angiotensin II(Angiotensin II) ELISA Kit

Alias
ANG II ELISA Kit, Angiotensin II ELISA Kit

Catalogue No.
EB0156

Size
48T/96T

Species
Bovine

UniProt ID
P01017

Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples

Detection Method
Competitive ELISA, Coated with antibody

Detection Wavelength
OD450

Reaction Duration
1 hours

Range
1.563-100pg/ml

Sensitivity
0.938pg/ml

Storage
2-8°C(Sealed), Don't cryopreserve.

Specificity
Specifically binds with ANGII , no obvious cross reaction with other analogues.

ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 HRP-Antigen(Concentrated, 100X) 30ul 60ul 2-8°C (Avoid Direct Light)
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antigen Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy

Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water

Assay Procedure Summary
Elisa实验原理图
  • Step 1: Wash the plate twice before adding the standard and sample.
  • Step 2: Add 100ul standard/sample/HRP-labeled antigen working solution into each well. Seal the plate and statically incubate for 45 minutes at 37°C. (Equilibrate the whole bottle of TMB for 30min at 37°C.)
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 3: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 4: Add 50ul stop solution. Read at 450nm immediately and calculate.

Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 2.512 2.64 2.563
1.562 1.885 1.961 1.923
3.125 1.533 1.595 1.564
6.25 1.319 1.373 1.346
12.5 1.003 1.043 1.023
25 0.711 0.74 0.725
50 0.465 0.483 0.474
100 0.263 0.273 0.268
EB0156 Standard Curve Image

Recovery

Add a certain amount of ANGII into the sample. Calculate the recovery by comparing the measured value with the expected amount of ANGII in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 89-96 92
EDTA plasma(n=10) 86-104 95
Heparin plasma(n=10) 86-104 97

Linearity

Dilute the sample with a certain amount of ANGII at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 95-105% 84-100% 81-89%
EDTA plasma(n=10) 92-97% 85-101% 81-97%
Heparin plasma(n=10) 87-97% 82-94% 88-100%

Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 3.21 13.11 48.67 3.23 12.69 49.98
Standard deviation 0.15 0.55 2.64 0.16 0.59 2.56
CV(%) 4.68 4.23 5.43 4.82 4.68 5.12

Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100